A new electron impact storage ion source has been designed for time of flight mass spectrometers with a high mass resolving power and high sensitivity. Because of the storage and compressing focus of the source, the overall performance of the instrument has been improved greatly with a mass resolving power up to 3 500. The conditions of the second order of the focus are given and the storage ability is proved. The competer simulations and experiment results are given, too.